EWA Release 40.8.1

Version 40.8.1 stable: Component Versions​

Version 40.8.1 stable of EWA was released to customers on 19th October 2026.

The compatible versions of the affected runtime components are listed below:

ComponentVersion
Chat Server26.8.274.0
Chat Service (Chat v2)0.14.0
Client Application26.10.83.0
Client Hub26.9.241.0
Data Warehouse Export Service26.9.222.0
EOC Integration26.9.222.0
Export Service26.9.231.0
Insights API26.9.223.0
Form Module API26.8.4
HP Link Patient Identity API26.9.231.0
Insights Web26.10.153.0
LiveView API26.9.222.0
Medical Unit Broker26.5.13.0
Migration and Seeding Tool26.9.222.0
Personnel Registration API26.8.1.0
Version Manager API26.9.72.0
Version Manager Client Installer26.9.78.0
EWA Multiplatform Client0.0.5385
EWA Multiplatform IIS0.0.5385

Version 40.8.1 stable: Upgrade Notes​

If you are upgrading from 40.8.0, you only need to upgrade the following two components for the fix in this release to take effect:

ComponentVersionFix
Client Application26.10.83.0The time picker only allowed 5-minute steps
Insights Web26.10.153.0The time picker only allowed 5-minute steps

The other components have no changes in 40.8.1 and can stay on their 40.8.0 versions.

Version 40.8.1 stable: New Features​

No new features were included in this release.

Version 40.8.1 stable: Fixes​

  1. The time picker only allowed 5-minute steps

The time picker only allowed 5-minute steps​

  • Ticket: 435672538337, 437225422061 | DevOps: 44384 | Reported by: Helse Nord, Helse Midt-Norge
  • HubSpot Title: Feil ved nøyaktig tid i Trombolyseskjema (STEMI); Journalnotat og Skjema: Ikke lenger mulig å velge 1-minutts intervall
  • Issue: The time picker in Forms (Time and Date+Time elements) and in History of illness only offered minutes in 5-minute steps (0, 5, 10 ... 55). A crew could not record an exact time such as 08:32, for example in the thrombolysis (STEMI) form, and an existing answer off the 5-minute grid opened with no minute highlighted.
  • Root cause: The minute column of the shared time picker was built from multiples of 5 only. The limit was introduced in EWA 37.0 together with the reworked, touch-scrollable picker.
  • Solution: The time picker in the Forms module lists every minute from 0 to 59 again, wrapping from 59 back to 0, and the change applies to both Forms and History of illness. Scrolling the column by touch and by mouse wheel works as before.

Affected components: Client Application (Windows / UWP), Insight Web

✅ Testing Focus Areas for 40.8.1​

🕒 Time picker in Forms and History of illness​

  • Open a form with a Time element, open the time picker and confirm the minute column lists every minute from 0 to 59, wrapping from 59 back to 0
  • Repeat on a Date+Time element, and on a time field in History of illness
  • Pick a minute that is not a multiple of 5, for example 7, and confirm the field shows it
  • Reopen the picker on that value and confirm minute 7 is highlighted
  • On a tablet, confirm the minute column can still be scrolled by touch, and on a PC by mouse wheel

Tested Medical Devices​

All versions of this release have been tested on the listed medical devices and corresponding software versions, ensuring compatibility and performance in the specified environments.

Corpuls​

  • Software versions: 4.4.4
  • SDK version: 4.4.0.0

Zoll​

  • Software version: 02.36.21.00
  • SDK version: 6.44.315