EWA Release 40.2.8
Version 40.2.8 stable: Component Versions
Version 40.2.8 stable of EWA was released to customers on April 13th 2026.
The compatible versions of the affected runtime components are listed below:
| Component | Version |
|---|---|
| Client Application | 26.4.41.0 |
Version 40.2.8 stable: New Features
No new features were included in this release.
Version 40.2.8 stable: Updated Features
No updated features were included in this release.
Version 40.2.8 stable: Fixes
EWA Client application crashes
- Ticket: 413997021430, 409788066016, 410087726315 | DevOps: 38594, 38264| Reported by: HSØ, HMN
- Issue: Users report that after the latest update the app crashes unpredictably—commonly while entering information in the old history of illness and when completing the cardiac arrest form—resulting in a full logout and requiring the user to sign in again. Technically, the UWP app crashes with
RPC_E_WRONG_THREAD (0x8001010E)when a background thread updates journal properties that trigger XAML binding operations on the wrong thread. - Solution: Wrapped the property assignment in DispatchHelper.DispatchToUiThread() to ensure all UI-related property changes happen on the correct thread.
- Affected components: Client Application
Tested Medical Devices
All versions of this release have been tested on the listed medical devices and corresponding software versions, ensuring compatibility and performance in the specified environments.
Corpuls
- Software versions: 4.3.2
- SDK version: 4.3.0.0
Zoll
- Software version: 02.36.21.00
- SDK version: 6.44.315